Blog with interesting cases and/or problems related to anesthesia with discussion based on best evidence in the literature.

Showing posts with label AICD. Show all posts
Showing posts with label AICD. Show all posts

March 29, 2016

Update-CIEDs



Recently I had a case where a 61 YO female underwent axillary mass excision. The patient had significant comorbidities, including non ischemic cardiomyopathy with EF estimated at 20%.  The patient had undergone multiple anesthetics in the past without problems.  The patient had a Boston Scientific AICD placed in 2011.

I have written on this topic in the past and noted that typically Boston Scientific AICDs, require auscultation to determine if magnet placement has discontinued the 'anti tachycardia therapy'.  In discussing this with the sales rep of the device, he stated, that as of 2006, all AICDs were programmed to turn off when  a magnet was placed over the device, and resume regular function when the magnet was removed.

Therefore, we elected to place defibrillation pads on the patient and turn off the anti tachy arrhythmia function of the AICD with a magnet and proceed to surgery.  The surgery was uneventful and the patient emerged from anesthesia after an uneventful anesthetic.

In the Fall APSF newsletter, the front page article featured information regarding CIEDs. It was emphasized that practitioners must understand what kind of device the patient has and how it is expected to function.   There are three main categories of CIEDs:  pacemaker, ICD, and devices that also perform resynchronization therapy (CRT).  This latter category is important because these are patients with significant cardiac dysfunction, and it is important to maintain the normal function of this device during the anesthetic. Fortunately, placing a magnet over this type of device will only inhibit the anti tachycardia therapy portion, and its pace making capabilities will function normally. However, if this patient is pacemaker dependent, either the device will need to be reprogrammed into a VOO mode, or, electrocautery will need to be limited.

In most cases, all pertinent information can be gained by interviewing the patient or having a chat with their cardiologist.  In some cases, however, information may not be available.  In some instances, a chest X-ray may be indicated that may provide information that can be used to determine what kind of device you have and what needs to be done if anything.  The clinical circumstances will dictate the necessity of a chest x ray which is almost guaranteed to delay a roll back to the OR.

For example, if surgery is distant from the generator (i.e. below the umbilicus), or if surgery is performed with a bipolar electrocautery unit, no changes in the device are likely needed beyond determining the indications for placement and verifying that it is functioning (i.e. the patient has had regular follow up with an EP clinic).

Patients who present to surgery with an ICD who also are pacemaker dependent present unique challenges.  Placing a magnet over the ICD will disable anti tachycardia pacing and defibrillation, but the pacemaker is unaffected and thus remains vulnerable to EMI interference from an electrocautery unit.  In this case, the surgeon will need to be notified that bipolar electrocautery is optimal to avoid interference with the pacemaker function if surgery is the in the chest or near the pacemaker generator.  If not, monopolar cautery can be used, but in short bursts to avoid prolonged interference. Alternatively, if available, the unit should be reprogrammed to address issues (i.e. disable the anti tachycardia and bradycardia therapy).


If an x ray is required to make a determination of what type of device the patient has, the following clues will be important.  In the above image, note there are THREE leads, a right atrial lead, a right ventricular lead and a coronary sinus lead.  Careful observation of the right ventricular lead allows notice of a thicker design than the other leads, which represents the shocking coil of an AICD.  The 3rd lead (or coronary sinus lead) suggests that this is for CRT for low ejection fraction.


In the above image, this patient came to surgery for emergency laparotomy.  She stated that she had a 'pacemaker' placed one month ago.  She was unsure of any of the details and family wasn't available. The above chest x ray indicates that this is not a pacemaker however, as there are no leads.   This is an implanted cardiac loop recorder.

Below, a dual chamber PM next to an AICD for comparison of difference.
Note that both the dual chamber PM and the AICD have a left atrial and left ventricular lead, but the left ventricular lead has a long radio opaque coil for electroshock delivery .

Note in this image the thick coils  to deliver electroshock in both the SVC and RV, plus an additional lead in the LV indicating the ability for resynchronization therapy.
In many patients with low EF, resynchronization therapy is utilized to improve ventricular function.  In most cases, these patients will have three leads, Atrial, and two ventricular.  However, in patients in chronic persistent atrial fibrillation, the electrophysiologist will only insert two ventricular leads and pace the ventricles only.  Therefore, only two leads are used and will be visible on chest X ray.  No coil will appear if no AICD is used.  Since patients who qualify for a biventricular pacemaker for resynchronization therapy have also traditionally also qualified for an AICD, it is unusual to see a biventricular pacemaker only.  However, recently, the indication for a biventricular pacemaker have changed and a larger number of patients are receiving these devices despite an EF of >40%. In the end, careful evaluation of the chest X ray can help clarify questions as to the type of device a patient has implanted.
Although devices implanted after about 2005, are very user friendly, patients may still present with older devices.  In this case, a thorough understanding of how they will respond to a magnet is important and therefore, it may be prudent to call a 24 hour tech support line.
Below are numbers as presented in the APSF write up in fall of 2013:

Medtronic---800-633-8766
St. Jude---800-722-3774
Boston Scientific---800-227-3422
Biotronik---800-547-0394
Note: Boston Scientific purchased 'Guidant'.


helpful links:
BJA -perioperative management of patients with cardiac implantable devices
HRS/ASA guidelines expert opinion
APSF article cited above
RSNA imaging article of cardiac devices





August 29, 2011

Boston Scientific AICD

A few days ago a patient came to the OR for a hip pinning. The patient had a biventricular pacer and AICD. Presumably this was for his significant heart failure (likely class III). Since the patient was unable to give a history, we referred to the chart for information. A cardiologist had seen the patient and recommended that the patient have a magnet placed over the pacer to turn off the tachyarrythmia function. It was noted that the patient had a Boston Scientific. In an earlier post on this issue I detailed how some pacers/defibrillators may respond in different ways to placement of a magnet. My colleague was doing the anesthetic so I suggested that he not place a magnet since EMI was not anticipated and bipolar was possible at a site (the hip) distant from the pacer leads.
Please see previous posts on this topic: pacemakers and magnet, AICD.

Boston Scientific has merged with Guidant, however, they still have pacemakers and AICDs floating around. In the end, it is really important to contact the sales rep for the device to understand how best to proceed if there is any thought that EMI will occur.

In the case of a Boston Scientific AICDs, the company has provided some guidance.

1. First determine if the unit can be turned off using a magnet. Place a magnet over the device and ausculate using a stethoscope. If you hear tones then you have two types: a long constant tone or a beeping tone. If you hear a long constant tone then the tachy Mode is programmed OFF. This means that the patient came to you with no automatic defibrillation capability. You may need to contact a rep to program on if the patient needs the availability of this function.
2. If you hear beeping tones then continue to listen for 30 sec. If the beeps CHANGE to a long constant tone then you can remove the magnet; you have successfully inhibited anti-tachy therapy. At the end of the case to turn the therapy back on, replace the magnet and listen until tones change to beeping; remove and therapy will be on.
3. If after placing the magnet you hear beeping tones but after 30 s there is no change, you must LEAVE the magnet in place to maintain inhibition (which you have acheived by placing the magnet). After the case, you can remove the magnet and therapy will be programmed ON.

Unfortunately, the article from Boston Scientific (dated March, 2009) has some slight contradictions. For example, it explains in the detailed desription that a change in tone from long to beeps or from beeps to long is equivalent and means that the device is programmed off. By replacing the magnet for 30 sec another change in tones (i.e. from long constant tone to beeps or from beeps to long constant) the device has been reprogrammed to on. For this reason, it is likely that it will be prudent to call a technician to better understand how to proceed.

There are two take home points from this post:
1) You CAN'T just place a magnet over any type of AICD and be 100% confident that you have disarmed the anti-tachy therapy.
2) Even if you know what model and brand you have, you probably should consult the sales rep by phone, and NOT assume the cardiologist knows what he/she is talking about. In our case placing a magnet over the pulse generator as recommended by the cardiologist may have: 1) done nothing at all as some Boston Scientific pulse generators do not reprogram with magnet placement or if the magnet was not left in place for at least 30 sec, 2) turned off therapy, but not have turned it back on at the end of the case when the magnet was removed, 3) turned off therapy and turned it back on at the end of the case when it was removed.
Finally, it should be noted that if you do turn off the anti-tachy therapy, please remember that you really should place defibrillator pads in position and be ready to shock the patient if the need should arise suddenly during the procedure.